R.J. Von Gutfeld, A.H. Nethercot Jr.
Physical Review Letters
A novel vacuum sample holder is described for use with an optical microscope. The sample stage temperature can be made to vary from ∼ 30 to 800 K. Electrical feedthroughs together with an optical window provide convenient features for use in a wide variety of experiments. Data for optically induced transverse thermoelectric voltages of a Pd film as a function of temperature are presented.
R.J. Von Gutfeld, A.H. Nethercot Jr.
Physical Review Letters
R.J. Von Gutfeld, F.Alan McDonald, et al.
Applied Physics Letters
R.J. Von Gutfeld, R.L. Melcher
Applied Physics Letters
R.J. Von Gutfeld
Applied Physics Letters